Futek Advanced Sensor Technology – Futek VCal™Documentation Rev. 1.8.0 07/26/05
©Futek Advanced Sensor Technology 2005
28
The final window under the Information Tabs Window
- is the Test Tab Window. This window provides you
some options for applying correction factors and
uncertainty values to your test to provide added
confidence in the accuracy of your results.
The Corrections section of this window is where
you can view the Linearization Coefficients and
Uncertainty Values for your calibration. Once you
have completed a test you can come to this tab
and click the Linearize Button to view these values.
Please note that Hysteresis Correction values will
only be calculated if Hysteresis was selected on the
Test Setup Tab under Test Conditions.
The Scaling section of this window is where you
can set Over Load and Over Range warning points,
as well as set Scaling Factors. You simply enter the
Rated Output of your sensor, load to the desired
calibration load, take a current output reading,
and then press Scale To Unit (more detail will be
given in the Instructions section). The number of
decimal points for pre-calibrated reference sensor
output can be entered in the Output Decimal Points
textbox.
Calibration Test Grid
The other half of the Reference Sensor interface
is
The C alibration Test Grid.
The Data Acquisition Channel is the channel you
have your Reference Sensor connected to.
The Test Grid is where you will perform your tests
and view your test results.
The Tare Button allows you to utilize offsets and
zero the unloaded output.
The Gross Button allows you to take an untared
reading with all offsets in place.